RESUMO

A hearing aid or a contralateral routing of signal device are options for unilateral cochlear implant listeners with limited hearing in the unimplanted ear; however, it is uncertain which device provides greater benefit beyond unilateral listening alone. Eighteen unilateral cochlear implant listeners participated in this prospective, within-participants, repeated measures study. Participants were tested with the cochlear implant alone, cochlear implant + hearing aid, and cochlear implant + contralateral routing of signal device configurations with a one-month take-home period between each in-person visit. Audiograms, speech perception in noise, and lateralization were evaluated. Subjective feedback was obtained via questionnaires. Marked improvement in speech in noise and non-implanted ear lateralization accuracy were observed with the addition of a contralateral hearing aid. There were no significant differences in speech recognition between listening configurations. However, the chronic device use questionnaires and the final device selection showed a clear preference for the hearing aid in spatial awareness and communication domains. Individuals with limited hearing in their unimplanted ears demonstrate significant improvement with the addition of a contralateral device. Subjective questionnaires somewhat contrast with clinic-based outcome measures, highlighting the delicate decision-making process involved in clinically advising one device or another to maximize communication benefits.


Assuntos
Implante Coclear , Implantes Cocleares , Auxiliares de Audição , Localização de Som , Percepção da Fala , Humanos , Estudos Prospectivos , Audição
2.
Cochlear Implants Int ; 24(5): 273-281, 2023 09.
Artigo em Inglês | MEDLINE | ID: mdl-37489512

RESUMO

OBJECTIVE: To investigate whether revision surgery with the same device results in a change in three key indicators of electrode positioning: scalar location, mean modiolar distance (M¯), and angular insertion depth (AID). METHODS: Retrospective analysis of a cochlear implant database at a university-based tertiary medical center. Intra-operative CT scans were obtained after initial and revision implantation. Electrode array (EA) position was calculated using auto-segmentation techniques. Initial and revision scalar location, M¯, and AID were compared. RESULTS: Mean change in M¯ for all ears was -0.07 mm (SD 0.24 mm; P = 0.16). The mean change in AID for all ears was -5° (SD 67°; P = 0.72). Three initial implantations with pre-curved EAs resulted in a translocation from Scala Tympani (ST) to Scala Vestibuli (SV). Two remained translocated after revision, while one was corrected when revised with a straight EA. An additional five translocations occurred after revision. CONCLUSIONS: In this study examining revision cochlear implantation from a single manufacturer, we demonstrated no significant change in key indicators of EA positioning, even when revising with a different style of electrode. However, the revision EA is not necessarily confined by the initial trajectory and there may be an increased risk of translocation.


Assuntos
Implante Coclear , Implantes Cocleares , Humanos , Estudos Retrospectivos , Cóclea , Reimplante
3.
Am J Audiol ; 32(2): 403-416, 2023 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-37249492

RESUMO

PURPOSE: This study investigated the relationship between the number of active electrodes, channel stimulation rate, and their interaction on speech recognition and sound quality measures while controlling for electrode placement. Cochlear implant (CI) recipients with precurved electrode arrays placed entirely within scala tympani and closer to the modiolus were hypothesized to be able to utilize more channels and possibly higher stimulation rates to achieve better speech recognition performance and sound quality ratings than recipients in previous studies. METHOD: Participants included seven postlingually deafened adult CI recipients with Advanced Bionics Mid-Scala electrode arrays confirmed to be entirely within scala tympani using postoperative computerized tomography. Twelve conditions were tested using four, eight, 12, and 16 electrodes and channel stimulation rates of 600 pulse per second (pps), 1,200 pps, and each participant's maximum allowable rate (1,245-4,800 pps). Measures of speech recognition and sound quality were acutely assessed. RESULTS: For the effect of channels, results showed no significant improvements beyond eight channels for all measures. For the effect of channel stimulation rate, results showed no significant improvements with higher rates, suggesting that 600 pps was sufficient for maximum speech recognition performance and sound quality ratings. However, across all conditions, there was a significant relationship between mean electrode-to-modiolus distance and all measures, suggesting that a lower mean electrode-to-modiolus distance was correlated with higher speech recognition scores and sound quality ratings. CONCLUSION: These findings suggest that even well-placed precurved electrode array recipients may not be able to take advantage of more than eight channels or higher channel stimulation rates (> 600 pps), but that closer electrode array placement to the modiolus correlates with better outcomes for these recipients.


Assuntos
Implante Coclear , Implantes Cocleares , Percepção da Fala , Adulto , Humanos , Percepção da Fala/fisiologia , Cóclea , Implante Coclear/métodos , Rampa do Tímpano/cirurgia
4.
J Acoust Soc Am ; 152(1): 67, 2022 07.
Artigo em Inglês | MEDLINE | ID: mdl-35931512

RESUMO

This study investigated the number of channels needed for maximum speech understanding and sound quality in 15 adult cochlear implant (CI) recipients with Advanced Bionics (AB) Mid-Scala electrode arrays completely within scala tympani. In experiment I, CI programs used a continuous interleaved sampling (CIS)-based strategy and 4-16 active electrodes. In experiment II, CI programs used an n-of-m strategy featuring 16 active electrodes with either 8- or 12-maxima. Speech understanding and sound quality measures were assessed. For CIS programs, participants demonstrated performance gains using up to 4-10 electrodes on speech measures and sound quality ratings. For n-of-m programs, there was no significant effect of maxima, suggesting 8-maxima is sufficient for this sample's maximum performance and sound quality. These results are largely consistent with previous studies using straight electrode arrays [e.g., Fishman, Shannon, and Slattery (1997). J. Speech Lang. Hear. Res. 40, 1201-1215; Friesen, Shannon, Baskent, and Wang (2001). J. Acoust. Soc. Am. 110, 1150-1163; Shannon, Cruz, and Galvin (2011). Audiol. Neurotol. 16, 113-123; Berg, Noble, Dawant, Dwyer, Labadie, and Gifford (2020). J. Acoust. Soc. Am. 147, 3646-3656] and in contrast with recent studies looking at cochlear precurved electrode arrays [e.g., Croghan, Duran, and Smith (2017). J. Acoust. Soc. Am. 142, EL537-EL543; Berg, Noble, Dawant, Dwuer, Labadie, and Gifford (2019b). J. Acoust. Soc. Am. 145, 1556-1564], which found continuous improvements up to 16 independent channels. These findings suggest that Mid-Scala electrode array recipients demonstrate similar channel independence to straight electrode arrays rather than other manufacturer's precurved electrode arrays.


Assuntos
Implante Coclear , Implantes Cocleares , Percepção da Fala , Adulto , Implante Coclear/métodos , Humanos , Rampa do Tímpano/cirurgia , Fala
5.
Otol Neurotol ; 43(7): 789-796, 2022 08 01.
Artigo em Inglês | MEDLINE | ID: mdl-35861647

RESUMO

OBJECTIVES: To evaluate the utility of intracochlear electrocochleography (ECochG) monitoring during cochlear implant (CI) surgery on postoperative hearing preservation. STUDY DESIGN: Prospective, randomized clinical trial. SETTING: Ten high-volume, tertiary care CI centers. PATIENTS: Adult patients with sensorineural hearing loss meeting the CI criteria who selected an Advanced Bionics CI. METHODS: Patients were randomized to CI surgery either with audible ECochG monitoring available to the surgeon during electrode insertion or without ECochG monitoring. Hearing preservation was determined by comparing preoperative unaided low-frequency (125-, 250-, and 500-Hz) pure-tone average (LF-PTA) to postoperative LF-PTA at CI activation. Pre- and post-CI computed tomography was used to determine electrode scalar location and electrode translocation. RESULTS: Eighty-five adult CI candidates were enrolled. The mean (standard deviation [SD]) unaided preoperative LF-PTA across the sample was 54 (17) dB HL. For the whole sample, hearing preservation was "good" (i.e., LF-PTA change 0-15 dB) in 34.5%, "fair" (i.e., LF-PTA change >15-29 dB) in 22.5%, and "poor" (i.e., LF-PTA change ≥30 dB) in 43%. For patients randomized to ECochG "on," mean (SD) LF-PTA change was 27 (20) dB compared with 27 (23) dB for patients randomized to ECochG "off" ( p = 0.89). Seven percent of patients, all of whom were randomized to ECochG off, showed electrode translocation from the scala tympani into the scala vestibuli. CONCLUSIONS: Although intracochlear ECochG during CI surgery has important prognostic utility, our data did not show significantly better hearing preservation in patients randomized to ECochG "on" compared with ECochG "off."


Assuntos
Implante Coclear , Implantes Cocleares , Adulto , Audiometria de Resposta Evocada/métodos , Cóclea/diagnóstico por imagem , Cóclea/cirurgia , Implante Coclear/métodos , Implantes Cocleares/efeitos adversos , Audição , Humanos , Estudos Prospectivos
6.
Am J Audiol ; 31(2): 380-391, 2022 Jun 02.
Artigo em Inglês | MEDLINE | ID: mdl-35549520

RESUMO

PURPOSE: Until recently, there has been little investigation on the effects of cochlear implantation on the transmission of acoustic stimuli through the middle-ear system. Recent studies have shown that cochlear implantation decreases low-frequency acoustic absorbance, consistent with a stiffer middle-ear system postsurgery. The objectives of this study are (a) to investigate the time course of changes in acoustic absorbance post-cochlear implantation in the implanted ear and (b) to compare changes in acoustic absorbance between implanted and nonimplanted ears over time. METHOD: Seventeen adult cochlear implant (CI) recipients within 6 months of device activation participated in this study. Wideband acoustic absorbance was measured in both ears at one to six different time points from pre-implantation up to 6-month postactivation. Analyses examined (a) changes in acoustic absorbance as compared to pre-implantation and (b) differences in acoustic absorbance between implanted and nonimplanted ears over time. RESULTS: Acoustic absorbance in the implanted ear decreased postsurgery for frequencies lower than 1.5 kHz and persisted through at least 6-month postactivation. We also observed that the spectral range of decreased acoustic absorbance in the implanted ear decreased with longer time postsurgery. Differences in acoustic absorbance between implanted and nonimplanted ears occurred over a broad spectral range at the activation time point and persisted through at least 3-month postactivation, though for a narrower spectral range at the later time point. CONCLUSIONS: Cochlear implantation increased middle-ear stiffness as indicated by decreased acoustic absorbance of low-frequency acoustic power. The findings of this study are consistent with those of previous studies and may have important implications toward understanding spatial hearing and programming of acoustic components for CI-combined electric and binaural acoustic stimulation patients.


Assuntos
Implante Coclear , Implantes Cocleares , Perda Auditiva Neurossensorial , Percepção da Fala , Acústica , Adulto , Audição , Humanos , Percepção da Fala/fisiologia
7.
Biomed Phys Eng Express ; 9(1)2022 11 23.
Artigo em Inglês | MEDLINE | ID: mdl-36594887

RESUMO

Objective. The cochlear implant is a neural prosthesis designed to directly stimulate auditory nerve fibers to induce the sensation of hearing in those experiencing severe-to-profound hearing loss. After surgical implantation, audiologists program the implant's external processor with settings intended to produce optimal hearing outcomes. The likelihood of achieving optimal outcomes increases when audiologists have access to tools that objectively present information related to the patient's own anatomy and surgical outcomes. This includes visualizations like the one presented here, termed the activation region overlap image, which is designed to decrease subjectivity when determining amounts of overlapping stimulation between implant electrodes.Approach. This visualization uses estimates of electric field strength to indicate spread of neural excitation due to each electrode. Unlike prior visualizations, this method explicitly defines regions of nerves receiving substantial stimulation from each electrode to help clinicians assess the presence of significant overlapping stimulation. A multi-reviewer study compared this and an existing technique on the consistency, efficiency, and optimality of plans generated from each method. Statistical significance was evaluated using the two-sided Wilcoxon rank sum test.Main results. The study showed statistically significant improvements in consistency (p < 10-12), efficiency (p < 10-15), and optimality (p < 10-5) when generating plans using the proposed method versus the existing method.Significance. This visualization addresses subjectivity in assessing overlapping stimulation between implant electrodes, which currently relies on reviewer estimates. The results of the evaluation indicate the provision of such objective information during programming sessions would likely benefit clinicians in making programming decisions.


Assuntos
Implante Coclear , Implantes Cocleares , Humanos , Implante Coclear/métodos , Eletrodos Implantados
8.
Artigo em Inglês | MEDLINE | ID: mdl-34790885

RESUMO

Disability is an important and often overlooked component of diversity. Individuals with disabilities bring a rare perspective to science, technology, engineering, mathematics, and medicine (STEMM) because of their unique experiences approaching complex issues related to health and disability, navigating the healthcare system, creatively solving problems unfamiliar to many individuals without disabilities, managing time and resources that are limited by physical or mental constraints, and advocating for themselves and others in the disabled community. Yet, individuals with disabilities are underrepresented in STEMM. Professional organizations can address this underrepresentation by recruiting individuals with disabilities for leadership opportunities, easing financial burdens, providing equal access, fostering peer-mentor groups, and establishing a culture of equity and inclusion spanning all facets of diversity. We are a group of deaf and hard-of-hearing (D/HH) engineers, scientists, and clinicians, most of whom are active in clinical practice and/or auditory research. We have worked within our professional societies to improve access and inclusion for D/HH individuals and others with disabilities. We describe how different models of disability inform our understanding of disability as a form of diversity. We address heterogeneity within disabled communities, including intersectionality between disability and other forms of diversity. We highlight how the Association for Research in Otolaryngology has supported our efforts to reduce ableism and promote access and inclusion for D/HH individuals. We also discuss future directions and challenges. The tools and approaches discussed here can be applied by other professional organizations to include individuals with all forms of diversity in STEMM.

9.
Trends Hear ; 25: 23312165211014139, 2021.
Artigo em Inglês | MEDLINE | ID: mdl-34027718

RESUMO

Individuals with bilateral cochlear implants (BiCIs) rely mostly on interaural level difference (ILD) cues to localize stationary sounds in the horizontal plane. Independent automatic gain control (AGC) in each device can distort this cue, resulting in poorer localization of stationary sound sources. However, little is known about how BiCI listeners perceive sound in motion. In this study, 12 BiCI listeners' spatial hearing abilities were assessed for both static and dynamic listening conditions when the sound processors were synchronized by applying the same compression gain to both devices as a means to better preserve the original ILD cues. Stimuli consisted of band-pass filtered (100-8000 Hz) Gaussian noise presented at various locations or panned over an array of loudspeakers. In the static listening condition, the distance between two sequentially presented stimuli was adaptively varied to arrive at the minimum audible angle, the smallest spatial separation at which the listener can correctly determine whether the second sound was to the left or right of the first. In the dynamic listening condition, participants identified if a single stimulus moved to the left or to the right. Velocity was held constant and the distance the stimulus traveled was adjusted using an adaptive procedure to determine the minimum audible movement angle. Median minimum audible angle decreased from 17.1° to 15.3° with the AGC synchronized. Median minimum audible movement angle decreased from 100° to 25.5°. These findings were statistically significant and support the hypothesis that synchronizing the AGC better preserves ILD cues and results in improved spatial hearing abilities. However, restoration of the ILD cue alone was not enough to bridge the large performance gap between BiCI listeners and normal-hearing listeners on these static and dynamic spatial hearing measures.


Assuntos
Implante Coclear , Implantes Cocleares , Localização de Som , Percepção da Fala , Percepção Auditiva , Humanos , Ruído/efeitos adversos
10.
J Acoust Soc Am ; 149(4): 2752, 2021 04.
Artigo em Inglês | MEDLINE | ID: mdl-33940865

RESUMO

This study investigated the number of channels available to cochlear implant (CI) recipients for maximum speech understanding and sound quality for lateral wall electrode arrays-which result in large electrode-to-modiolus distances-featuring the greatest inter-electrode distances (2.1-2.4 mm), the longest active lengths (23.1-26.4 mm), and the fewest number of electrodes commercially available. Participants included ten post-lingually deafened adult CI recipients with MED-EL electrode arrays (FLEX28 and STANDARD) entirely within scala tympani. Electrode placement and scalar location were determined using computerized tomography. The number of channels was varied from 4 to 12 with equal spatial distribution across the array. A continuous interleaved sampling-based strategy was used. Speech recognition, sound quality ratings, and a closed-set vowel recognition task were measured acutely for each electrode condition. Participants did not demonstrate statistically significant differences beyond eight channels at the group level for almost all measures. However, several listeners showed considerable improvements from 8 to 12 channels for speech and sound quality measures. These results suggest that channel interaction caused by the greater electrode-to-modiolus distances of straight electrode arrays could be partially compensated for by a large inter-electrode distance or spacing.


Assuntos
Implante Coclear , Implantes Cocleares , Percepção da Fala , Adulto , Humanos , Rampa do Tímpano/cirurgia , Fala
11.
J Acoust Soc Am ; 147(5): 3646, 2020 05.
Artigo em Inglês | MEDLINE | ID: mdl-32486813

RESUMO

This study investigated the effects of cochlear implant (CI) electrode array type and scalar location on the number of channels available to CI recipients for maximum speech understanding and sound quality. Eighteen post-lingually deafened adult CI recipients participated, including 11 recipients with straight electrode arrays entirely in scala tympani and 7 recipients with translocated precurved electrode arrays. Computerized tomography was used to determine electrode placement and scalar location. In each condition, the number of channels varied from 4 to 22 with equal spatial distribution across the array. Speech recognition (monosyllables, sentences in quiet and in noise), subjective speech sound quality, and closed-set auditory tasks (vowels, consonants, and spectral modulation detection) were measured acutely. Recipients with well-placed straight electrode arrays and translocated precurved electrode arrays performed similarly, demonstrating asymptotic speech recognition scores with 8-10 channels, consistent with the classic literature. This finding contrasts with recent work [Berg, Noble, Dawant, Dwyer, Labadie, and Gifford. (2019). J. Acoust. Soc. Am. 145, 1556-1564] that found precurved electrode arrays well-placed in scala tympani demonstrate continuous performance gains beyond 8-10 channels. Given these results, straight and translocated precurved electrode arrays are theorized to have less channel independence secondary to their placement farther away from neural targets.


Assuntos
Implante Coclear , Implantes Cocleares , Percepção da Fala , Rampa do Tímpano/cirurgia , Fala
12.
J Am Acad Audiol ; 31(8): 578-589, 2020 09.
Artigo em Inglês | MEDLINE | ID: mdl-32340055

RESUMO

BACKGROUND: Microphone location has been shown to influence speech recognition with a microphone placed at the entrance to the ear canal yielding higher levels of speech recognition than top-of-the-pinna placement. Although this work is currently influencing cochlear implant programming practices, prior studies were completed with previous-generation microphone and sound processor technology. Consequently, the applicability of prior studies to current clinical practice is unclear. PURPOSE: To investigate how microphone location (e.g., at the entrance to the ear canal, at the top of the pinna), speech-source location, and configuration (e.g., omnidirectional, directional) influence speech recognition for adult CI recipients with the latest in sound processor technology. RESEARCH DESIGN: Single-center prospective study using a within-subjects, repeated-measures design. STUDY SAMPLE: Eleven experienced adult Advanced Bionics cochlear implant recipients (five bilateral, six bimodal) using a Naída CI Q90 sound processor were recruited for this study. DATA COLLECTION AND ANALYSIS: Sentences were presented from a single loudspeaker at 65 dBA for source azimuths of 0°, 90°, or 270° with semidiffuse noise originating from the remaining loudspeakers in the R-SPACE array. Individualized signal-to-noise ratios were determined to obtain 50% correct in the unilateral cochlear implant condition with the signal at 0°. Performance was compared across the following microphone sources: T-Mic 2, integrated processor microphone (formerly behind-the-ear mic), processor microphone + T-Mic 2, and two types of beamforming: monaural, adaptive beamforming (UltraZoom) and binaural beamforming (StereoZoom). Repeated-measures analyses were completed for both speech recognition and microphone output for each microphone location and configuration as well as sound source location. A two-way analysis of variance using mic and azimuth as the independent variables and output for pink noise as the dependent variable was used to characterize the acoustic output characteristics of each microphone source. RESULTS: No significant differences in speech recognition across omnidirectional mic location at any source azimuth or listening condition were observed. Secondary findings were (1) omnidirectional microphone configurations afforded significantly higher speech recognition for conditions in which speech was directed to ± 90° (when compared with directional microphone configurations), (2) omnidirectional microphone output was significantly greater when the signal was presented off-axis, and (3) processor microphone output was significantly greater than T-Mic 2 when the sound originated from 0°, which contributed to better aided detection at 2 and 6 kHz with the processor microphone in this group. CONCLUSIONS: Unlike previous-generation microphones, we found no statistically significant effect of microphone location on speech recognition in noise from any source azimuth. Directional microphones significantly improved speech recognition in the most difficult listening environments.


Assuntos
Implante Coclear , Implantes Cocleares , Percepção da Fala , Adulto , Humanos , Estudos Prospectivos , Fala
13.
Acta Otolaryngol ; 140(3): 206-211, 2020 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-31859576

RESUMO

Background: Imaging of cochlear implant (CI) electrode arrays (EAs) consists of intraoperative fluoroscopy to rule out tip fold-over and/or post-operative computerized tomography (CT) if concern exists regarding extrusion or misplacement of the EA. Intraoperative CT (iCT) can satisfy these current needs and enables specification of final intracochlear position.Aims/objectives: To describe iCT scanning of CI recipients at an academic medical center.Materials and methods: iCT was used to scan CI recipients within the operating room before recovering from general anesthesia.Results: In fiscal year 2019, 301 CI were placed (83 children, 218 adult). One hundred, seventy-five iCTs were performed (58% of total CIs) of which 52 were children (63% of pediatric CIs) and 123 were adult (57% of adult CIs). Of 7 CI surgeons, use of iCT ranged from 14% to 100% (mean 60%). Four tip fold-overs were identified and corrected intraoperatively. Surgeons reported using the images to improve technique (i.e. pulling back precurved EAs to improve perimodiolar positioning).Conclusion and significance: The current standard of care for CI is to insert EAs without feedback as to final location. iCT provides surgeons with rapid post-insertion feedback which allows detection and correction of suboptimally placed EAs as well as refinement of surgical technique.


Assuntos
Cóclea/diagnóstico por imagem , Implante Coclear/normas , Implantes Cocleares , Tomografia Computadorizada por Raios X , Adulto , Cadáver , Criança , Implante Coclear/métodos , Humanos , Cuidados Intraoperatórios , Controle de Qualidade , Padrão de Cuidado
14.
Otol Neurotol ; 41(1): 33-38, 2020 01.
Artigo em Inglês | MEDLINE | ID: mdl-31746820

RESUMO

OBJECTIVES: Describe audiologic outcomes in hearing preservation cochlear implantation (CI) using a precurved electrode array inserted using an external sheath and evaluate association of electrode positioning and preservation of residual hearing. STUDY DESIGN: Retrospective review. SETTING: Tertiary otologic center. PATIENTS: Twenty-four adult patients who underwent hearing preservation CI with precurved electrode array. INTERVENTIONS: CI, intraoperative computed tomography (CT) OUTCOME MEASURES:: Audiologic measures (consonant-nucleus-consonant [CNC] words, AzBio sentences, low-frequency pure tone averages [LFPTA]) and electrode location (scalar location, electrode-to-modiolus distance ((Equation is included in full-text article.)), angular insertion depth). RESULTS: Twenty-four adults with less than 80 dB LFPTA with a precurved electrode array inserted using an external sheath; 16 underwent intraoperative CT. LFPTA was 58.5 dB HL preoperatively, with a 17.3 dB threshold shift at CI activation (p = 0.005). CNC word scores improved from 6% preoperatively to 64% at 6 months postoperatively (p < 0.0001). There was one scalar translocation and no tip fold-overs. The average angular insertion depth was 388.2 degrees, and the average (Equation is included in full-text article.)across all electrodes was 0.36 mm. Multivariate regression revealed a significant correlation between CNC scores at 6 months and angular insertion depth (p = 0.0122; r = 0.45, adjusted r = 0.35). Change in LFPTA was not significantly associated with angular insertion depth or (Equation is included in full-text article.). CONCLUSIONS: A low rate of translocation allows a precurved electrode array inserted using an external sheath to maintain hearing preservation rates comparable to straight electrode arrays. With scala tympani insertion, angular insertion depth is a positive marker of improved speech performance postoperatively but may be a confounder variable based on individual cochlear size.


Assuntos
Implante Coclear/instrumentação , Implante Coclear/métodos , Implantes Cocleares , Audição , Resultado do Tratamento , Adulto , Feminino , Audição/fisiologia , Humanos , Masculino , Pessoa de Meia-Idade , Estudos Retrospectivos
15.
Otol Neurotol ; 40(9): 1160-1166, 2019 10.
Artigo em Inglês | MEDLINE | ID: mdl-31469799

RESUMO

OBJECTIVE: Characterize differences in adult cochlear implant outcomes and programming parameters for a straight (CI422/522) and a precurved (CI532) electrode array. SETTING: Cochlear implant (CI) program at a tertiary otologic center. PATIENTS: Fifty-eight adults were included in the study; 29 were implanted with CI422 or CI522 and 29 were implanted with CI532. Each CI532 recipient was matched to a CI422/522 recipient in terms of age and preoperative hearing thresholds for comparison purposes. MAIN OUTCOME MEASURES: Consonant-Nucleus-Consonant (CNC) words, AzBio sentences, residual audiometric thresholds, and Speech Spatial Qualities (SSQ) questionnaire collected 6 months postoperatively were used to characterize outcomes. Pulse duration, maxima, impedances, and overall charge measurements were used to characterize programming parameters. RESULTS: Postoperative unaided low frequency pure-tone average (LFPTA) was significantly better for the CI532 group. CNC scores were significantly better for the CI532 group. Impedances and pulse duration were significantly lower for the CI532 group, but there was no difference in overall charge between the groups. CONCLUSION: The CI532 group showed either similar or statistically superior results on all measures when compared with the CI422/522 suggesting that the CI532 electrode may be an advantageous substitute for the CI522.


Assuntos
Implantes Cocleares , Perda Auditiva Neurossensorial/cirurgia , Resultado do Tratamento , Adulto , Implante Coclear/métodos , Estudos de Coortes , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Período Pós-Operatório , Estudos Retrospectivos , Percepção da Fala
16.
Am J Audiol ; 28(3): 686-696, 2019 Sep 13.
Artigo em Inglês | MEDLINE | ID: mdl-31430174

RESUMO

Purpose There is a growing body of literature that suggests a linkage between impaired auditory function, increased listening effort, and fatigue in children and adults with hearing loss. Research suggests this linkage may be associated with hearing loss-related variations in diurnal cortisol levels. Here, we examine variations in cortisol profiles between young adults with and without severe sensorineural hearing loss and examine associations between cortisol and subjective measures of listening effort and fatigue. Method This study used a repeated-measures, matched-pair design. Two groups (n = 8 per group) of adults enrolled in audiology programs participated, 1 group of adults with hearing loss (AHL) and 1 matched control group without hearing loss. Salivary cortisol samples were collected at 7 time points over a 2-week period and used to quantify physiological stress. Subjective measures of listening effort, stress, and fatigue were also collected to investigate relationships between cortisol levels, perceived stress, and fatigue. Results Subjective ratings revealed that AHL required significantly more effort and concentration on typical auditory tasks than the control group. Likewise, complaints of listening-related fatigue were more frequent and more of a problem in everyday life for AHL compared to the control group. There was a significant association between subjective ratings of listening effort and listening-related fatigue for our AHL, but not for the control group. In contrast, there was no significant difference in cortisol measures between groups, nor were there significant associations between cortisol and any subjective measure. Conclusions Young AHL experience more effortful listening than their normal hearing peers. This increased effort is associated with increased reports of listening-related fatigue. However, diurnal cortisol profiles were not significantly different between groups nor were they associated with these perceived differences.


Assuntos
Ritmo Circadiano , Surdez/reabilitação , Fadiga/metabolismo , Hidrocortisona/metabolismo , Estresse Psicológico/metabolismo , Adulto , Estudos de Casos e Controles , Implantes Cocleares , Fadiga/psicologia , Feminino , Humanos , Masculino , Projetos Piloto , Estresse Psicológico/psicologia , Adulto Jovem
17.
Otol Neurotol ; 40(5): e503-e510, 2019 06.
Artigo em Inglês | MEDLINE | ID: mdl-31083085

RESUMO

HYPOTHESIS: Electrocochleography (ECochG) recorded during cochlear implant (CI) insertion from the apical electrode in conjunction with postinsertion ECochG can identify electrophysiologic differences that exist between groups with and without a translocation of the array from the scala tympani (ST) into the scala vestibuli (SV). BACKGROUND: Translocation of the CI electrode from ST into SV can limit performance postoperatively. ECochG markers of trauma may be able to aid in the ability to detect electrode array-induced trauma/scalar translocation intraoperatively. METHODS: Twenty-one adult CI patients were included. Subjects were postoperatively parsed into two groups based on analysis of postoperative imaging: 1) ST (n = 14) insertion; 2) SV (n = 7) insertion, indicating translocation of the electrode. The ECochG response elicited from a 500 Hz acoustic stimulus was recorded from the lead electrode during insertion when the distal electrode marker was at the round window, and was compared to the response recorded from a basal electrode (e13) after complete insertion. RESULTS: No statistically significant change in mean ECochG magnitude was found in either group between recording intervals. There was a mean loss of preoperative pure-tone average of 52% for the nontranslocation group and 94% for the translocation group. CONCLUSIONS: Intraoperative intracochlear ECochG through the CI array provides a unique opportunity to explore the impact of the CI electrode on the inner ear. Specifically, a translocation of the array from ST to SV does not seem to change the biomechanics of the cochlear region that lies basal to the area of translocation in the acute period.


Assuntos
Cóclea/cirurgia , Implante Coclear/métodos , Implantes Cocleares , Eletrodos , Adulto , Audiometria de Resposta Evocada , Audiometria de Tons Puros , Fenômenos Biomecânicos , Cóclea/diagnóstico por imagem , Humanos , Monitorização Intraoperatória , Estudos Prospectivos , Rampa do Tímpano , Rampa do Vestíbulo , Tomografia Computadorizada por Raios X , Resultado do Tratamento
18.
J Acoust Soc Am ; 145(3): 1556, 2019 03.
Artigo em Inglês | MEDLINE | ID: mdl-31067952

RESUMO

This study investigated the number of channels needed for maximum speech understanding and sound quality in 30 adult cochlear implant (CI) recipients with perimodiolar electrode arrays verified via imaging to be completely within scala tympani (ST). Performance was assessed using a continuous interleaved sampling (CIS) strategy with 4, 8, 10, and 16 channels and n-of-m with 16 maxima. Listeners were administered auditory tasks of speech understanding [monosyllables, sentences (quiet and +5 dB signal-to-noise ratio, SNR), vowels, consonants], spectral modulation detection, as well as subjective estimates of sound quality. Results were as follows: (1) significant performance gains were observed for speech in quiet (monosyllables and sentences) with 16- as compared to 8-channel CIS, (2) 16 channels in a 16-of-m strategy yielded significantly higher outcomes than 16-channel CIS for sentences in noise (percent correct and subjective sound quality) and spectral modulation detection, (3) 16 channels in a 16-of-m strategy yielded significantly higher outcomes as compared to 8- and 10-channel CIS for monosyllables, sentences (quiet and noise), consonants, spectral modulation detection, and subjective sound quality, (4) 16 versus 8 maxima yielded significantly higher speech recognition for monosyllables and sentences in noise using an n-of-m strategy, and (5) the degree of benefit afforded by 16 versus 8 maxima was inversely correlated with mean electrode-to-modiolus distance. These data demonstrate greater channel independence with perimodiolar electrode arrays as compared to previous studies with straight electrodes and warrant further investigation of the minimum number of maxima and number of channels needed for maximum auditory outcomes.


Assuntos
Implantes Cocleares/normas , Interface para o Reconhecimento da Fala , Fala , Adulto , Idoso , Idoso de 80 Anos ou mais , Eletrodos/normas , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Fonética
19.
Ear Hear ; 40(6): 1478-1480, 2019.
Artigo em Inglês | MEDLINE | ID: mdl-31033635

RESUMO

OBJECTIVES: The Quick Spectral Modulation Detection (QSMD) test provides a quick and clinically implementable spectral resolution estimate for cochlear implant (CI) users. However, the original QSMD software (QSMD(MySound)) has technical and usability limitations that prevent widespread distribution and implementation. In this article, we introduce a new software package EasyQSMD, which is freely available software with the goal of both simplifying and standardizing spectral resolution measurements. DESIGN: QSMD was measured for 20 CI users using both software packages. RESULTS: No differences between the two software packages were detected, and based on the 95% confidence interval of the difference between tests, the difference between the tests is expected to be <2% points. The average test duration was under 4 minutes. CONCLUSIONS: EasyQSMD is considered functionally equivalent to QSMD(MySound) providing a clinically feasible and quick estimate of spectral resolution for CI users.


Assuntos
Percepção Auditiva , Implantes Cocleares , Software , Implante Coclear , Perda Auditiva/reabilitação , Humanos , Reprodutibilidade dos Testes , Detecção de Sinal Psicológico
20.
J Am Acad Audiol ; 30(3): 235-242, 2019 03.
Artigo em Inglês | MEDLINE | ID: mdl-30461413

RESUMO

BACKGROUND: Bilateral cochlear implantation is the standard of care for individuals with moderate sloping-to-profound sensorineural hearing loss who do not receive benefit from appropriately fit hearing aids. Because of financial, insurance, or medical reasons, some unilateral cochlear implant (CI) recipients are unable to obtain a second CI. Here, we evaluated the first clinically available solution for individuals who have been unilaterally implanted and who do not or cannot use technology (e.g., hearing aid or CI) on the non-implanted ear. PURPOSE: We aimed to investigate how the addition of a contralateral routing of signal (CROS) device could provide objective and/or subjective benefit to adult CI recipients with moderate-to-profound hearing loss in the non-implanted ear. RESEARCH DESIGN: Single-center prospective study using a within-subjects repeated-measures design. STUDY SAMPLE: Participants included ten experienced unilateral CI recipients with severe-to-profound (n = 9) or moderate-to-profound (n = 1) sensorineural hearing loss in the non-implanted ear. At the time of study enrollment, participants did not use any technology on the non-implanted ear. No other exclusion criteria were used. INTERVENTION: Individuals were tested with and without a CROS device worn on the non-implanted ear. DATA COLLECTION AND ANALYSIS: We obtained measures of speech understanding in quiet (50 and 65 dBA) and in noise (+5-dB signal-to-noise ratio with a 65-dBA speech signal) both with and without the CROS device in an acute listening condition. Subjective benefit was assessed via the Speech, Spatial and Qualities 12-item questionnaire before CROS fitting and after two weeks of continuous use. A mixed-model, repeated-measures analysis of variance was completed with three talker locations and three presentation levels included as within-subjects factors and the presence or absence of a CROS device as a between-subjects factor. RESULTS: There was an 11% improvement in speech understanding in noise with the addition of the CROS device when speech was located at 0° azimuth. Subjective benefit in the speech domain of the SSQ was also observed. CONCLUSIONS: Use of CROS provided both subjective and objective speech recognition benefit for unilateral CI recipients who do not have access to bilateral cochlear implantation.


Assuntos
Implantes Cocleares , Auxiliares de Audição , Perda Auditiva Neurossensorial/reabilitação , Satisfação do Paciente , Localização de Som/fisiologia , Percepção da Fala/fisiologia , Seguimentos , Perda Auditiva Neurossensorial/fisiopatologia , Humanos , Ruído , Estudos Prospectivos , Inquéritos e Questionários
SELEÇÃO DE REFERÊNCIAS